ORIGINAL ARTICLE |
|
Year : 2022 | Volume
: 9
| Issue : 1 | Page : 14-23 |
|
Automatic speech processing software – New sensitive tool for the assessment of nasality: A preliminary study
K. S. Girish1, M. Pushpavathi2, Ajish K. Abraham3, C. M. Vikram4
1 Department of Speech Language Pathology, JSS Institute of Speech and Hearing, Kelageri, Dharwad, India 2 Director, All India Institute of Speech and Hearing, Mysuru, India 3 Department of Electronics and Acoustics, All India Institute of Speech and Hearing, India 4 Department of Speech and Hearing Science, Arizona State University, USA
Date of Submission | 11-Jun-2021 |
Date of Acceptance | 18-Sep-2021 |
Date of Web Publication | 01-Jan-2022 |
Correspondence Address: Mr. K. S. Girish JSS Institute of Speech and Hearing, Kelageri, Dharwad - 580 007, Karnataka India
 Source of Support: None, Conflict of Interest: None  | Check |
DOI: 10.4103/jclpca.jclpca_22_21
Introduction: Automatic speech processing (ASP) software is a nasality assessment tool. ASP studies focusing on investigating sentences to find nasality and correlating ASP scores with other objective assessment scores measuring nasality are scarce. Hence, the present study aimed at comparing the nasalance values of the ASP software with the nasometer in typically developing children (TDC) and children with repaired cleft palate (RCP) across different stimuli. Methods: Participants included 30 Kannada speaking TDC and 10 children with RCP (9–12 years). Speech stimuli (oral, nasal, and oronasal sentences) were recorded and the values were obtained from the ASP software as well as the nasometer. The following statistical tests were applied: mixed ANOVA, repeated measures ANOVA, paired samples t-test, independent samples t-test and Pearson's correlation. Results: Like nasometer, the nasalance values of ASP software were high for the nasal sentences followed by the oronasal sentences and the oral sentences, for both the populations. Higher nasalance values were found for children with RCP than for TDC across all the stimuli. Significant differences were found in nasalance values between the instruments in oral and oronasal sentences in TDC and nasal sentences and oronasal sentences in RCP. The nasalance values across the stimuli between nasometer and ASP software in both the groups showed no significant correlations. Conclusions: ASP software was successful in identifying nasalance in TDC and children with RCP. However, a major issue needs to be addressed concerning the dynamic range of the software and it has to be validated on a large number of populations.
Keywords: Automatic speech processing software, cleft lip and palate, hypernasality, nasometer, repaired cleft palate
How to cite this article: Girish KS, Pushpavathi M, Abraham AK, Vikram CM. Automatic speech processing software – New sensitive tool for the assessment of nasality: A preliminary study. J Cleft Lip Palate Craniofac Anomal 2022;9:14-23 |
How to cite this URL: Girish KS, Pushpavathi M, Abraham AK, Vikram CM. Automatic speech processing software – New sensitive tool for the assessment of nasality: A preliminary study. J Cleft Lip Palate Craniofac Anomal [serial online] 2022 [cited 2023 Jun 6];9:14-23. Available from: https://www.jclpca.org/text.asp?2022/9/1/14/333639 |
Introduction | |  |
Hypernasality is the abnormal coupling of oral and nasal cavities which results in excessive nasality on vowels and vocalic consonants.[1] Along with articulation errors and nasal air emission, hypernasality is one of the major speech errors affecting the intelligibility of speech in children with cleft palate.[2] Assessment of hypernasality can be done subjectively as well as objectively.
Several test materials and protocols are developed to perceptually assess hypernasality in children with cleft palate.[3],[4],[5],[6],[7],[8] However, this method carries inherent limitations, hence, subjective assessments are augmented by objective assessments to overcome these limitations. Instruments universally used for the measurement of nasality are nasometer, nasal view, and nasal visualization system. Among these, nasometer (Kay Elemetrics, USA) is the widely used objective assessment tool for the measurement of nasality for both clinical and research purposes.[9] It measures the nasalance values by separately measuring the nasal sound pressure level and oral sound pressure level, i.e., “nasalance = nasal energy/(nasal energy + oral energy) ×100.”[10] However, nasalance varies across various factors such as language, age, gender, cleft palate, and stimuli.[1],[11]
Over decades, nasometry studies have shown higher nasalance values for individuals with a history of the cleft palate than noncleft individuals.[12],[13] Across the stimuli, studies have revealed higher nasalance values for nasal stimuli, followed by oronasal and oral stimuli in noncleft individuals.[14],[15],[16],[17] Furthermore, the same trend was followed across the stimuli, in individuals with cleft lip and palate (CLP).[12],[13] Although nasometer has good sensitivity and specificity, there are some limitations which include expensive, inability to record the original signals for further analysis, nonportability, sanitization after each use, time-consuming test procedures, refusal to wear headgear by children, and the requirement of skilled transducer placement.[2],[18],[19]
To overcome the limitations, speech processing-based techniques have been recently proposed to analyze nasality in the cleft palate population. These methods analyze the acoustic characteristics of nasalization from speech signals and provide values. Many studies have been conducted using speech processing techniques in the assessment of nasality in noncleft individuals as well as in the cleft palate population.[20],[21] The authors have investigated automatic detection of nasality in noncleft and clinical populations using acoustic analysis and the results showed a good accuracy level in detecting nasality in vowels[22] and syllables/words.[23],[24] Most of the automatic speech processing (ASP) studies are done on vowels and words, but studies have rarely used sentences.[25] Furthermore, very few studies correlated nasality assessing ASP tools with perceptual or objective assessment tool assessments to check the reliability.[26],[27],[28]
It is clear that the automatic evaluation using the ASP software is a valuable means for research and clinical purpose to quantify nasality in children with CLP, but there are no commercially made instruments available using the ASP system. Hence, to overcome the limitations of the presently available objective assessment tools, ASP software was developed by speech scientists.[29] It is diagnostic computer software that works on the principle of ASP. In this system, the acoustic models for the nasals and oral sounds are developed. The likelihood scores computed for the nasal class are used as acoustic correlates of hypernasality. It is intended to use for the assessment of the severity of hypernasality in children with a history of cleft palate in the age range of 6–12 years. Scoring will be provided in the form of continuous numbers from “0” to “100.” However, the reliability of the software needs to be checked.
With the increasing incidence of cleft palate population in India (1.09–1.4 in 1000 live births),[30],[31] which is scattered over a wide region, it demands the need to have an objective assessment tool that is portable, aiding the SLP to reach clients in remote areas. In addition, it is relatively cost-effective when compared to nasometer, nasal view, and NVS. This software is compatible with most of the “windows” versions and can be easily installed on any laptop or computer. This software requires a simple dynamic microphone connected to the computer/laptop for recording and can be operated by speech professionals and other qualified professionals as well. Whereas, other available instruments require separate hardware and software of their own. It is also equipped with headgear with a sound separator plate, which small children refuse to wear.[19]
Furthermore, very few ASP studies have focused on investigating the use of sentences to measure nasality. Objective evaluation of hypernasality using sentences in individuals with CLP is important as the listener reliability for the perceptual rating of nasality is higher for sentences than for words and isolated vowels.[32] Comparison of nasalance values of the ASP software and nasometer helps in checking the effectiveness of the ASP software in identifying the nasalance. Furthermore, as a result of this preliminary study, we get to know the drawbacks of the ASP software, if any, enabling us to improvise the software accordingly. ASP software can be used across various medical and rehabilitation centers if proven effective. This will help in reaching a greater number of children with a history of CLP in turn helping in the early identification, diagnosis, and management of children with CLP. Hence, the present study aims at comparing the nasalance values of the ASP software and nasalance values of the nasometer in typically developing children (TDC) and children with repaired cleft palate (RCP) across different stimuli. The objectives are (1) comparison of nasalance values across the stimuli using nasometer and ASP software in TDC, (2) comparison of nasalance values across the stimuli using nasometer and ASP software in children with RCP, (3) comparison of nasalance values across the stimuli between the groups (TDC vs. RCP) in both nasometer and ASP software, (4) comparison of nasalance values between the instruments (nasometer and ASP software) across the stimuli in both the groups, and (5) correlation of nasalance values across the stimuli between nasometer and ASP software in both the groups.
Methods | |  |
Participants
Participants included 30 TDC and 10 nonsyndromic children with RCP (cleft palate alone – 7, CLP – 3) between 9 and 12 years of age. All the participants had Kannada as their native language. All the TDC were screened by the investigator, whereas all the children with RCP were tested by ENT, plastic surgeon, clinical psychologist, orthodontist, and SLP (investigator) at the “Unit for Structural Oro-Facial Anomalies” at the investigator's institute to rule out children with associated problems such as hearing loss, intellectual disability, language disorder, and nasal pathologies. Furthermore, the investigator perceptually screened all the children with RCP, and only those who had hypernasality were considered for this study. All the children with RCP had moderate-to-severe hypernasality. All the ethical guidelines were followed and written consent was taken from parents/caregivers of children with RCP and TDC.[33] Participants' details are provided in [Table 1]. | Table 1: Participants' details in typically developing children and repaired cleft palate
Click here to view |
Research design
The study involved comparing the nasalance values of children with RCP and TDC. Hence, the standard group comparison was employed.
Materials
The speech stimuli included five oral sentences, five nasal sentences, and five oronasal sentences [Appendix 1].[14] Oral sentences are loaded with high-pressure oral consonants, nasal sentences with nasal consonants, and oronasal sentences with both oral and nasal consonants.
Procedure
The present study included two objective test procedures. Both the test procedures were modeled by the investigator to each participant and were asked to follow the same. The test procedures are as follows:
Nasometry
Before recording, the nasometer (Model 6450, PENTAX Medical) was calibrated according to the manufacturer's instructions. Each participant was comfortably seated on a chair and the nasometer headgear was firmly placed against the upper lip perpendicular to the facial plane, as shown in [Figure 1]. | Figure 1: A child seated for the recording of speech samples in the nasometer
Click here to view |
Participants were instructed to carefully listen to the stimuli and repeat once clearly, in their comfortable pitch and loudness. The responses were recorded using the nasometer application and saved in “nsp” format for further analysis. In this application, the cursors on the screen were set from onset to the offset of the stimulus and the nasalance values were noted, as shown in [Figure 2]. | Figure 2: Stimulus selection and extraction of nasalance values for an oral stimulus (left) and a nasal stimulus (right)
Click here to view |
Automatic speech processing software
Each participant was comfortably seated on a chair in a sound-treated room. The participants were instructed to listen carefully and repeat the stimulus once after the investigator, in their comfortable pitch and loudness. The responses were audio recorded by placing a super-cardioid vocal dynamic microphone (Mipro MM-107) at 4 cm[34] away from the child's mouth using ASP software on a computer. The recorded speech samples were saved in “wav” format on the computer. For proof, the entire procedure was recorded using a Sony digital video camera recorder, as shown in [Figure 3]. | Figure 3: A child seated for the recording of speech samples in the automatic speech processing system
Click here to view |
The saved speech samples were analyzed by the investigator within the ASP software. The speech sample was loaded and the required stimulus was segmented on the screen from onset to the offset of the signal and the nasalance values were computed [as shown in [Figure 4]]. From the segmented speech signal, Mel-frequency cepstral coefficients (MFCCs) are computed. The MFCC features are given as input to the Gaussian mixture model (GMM)-based oral and nasal classifier. The posterior probability scores computed for the nasal class are used for the nasality score computation. In this work, the pretrained GMM model from an ASP study is used.[35] | Figure 4: Computation of nasalance values of an oral stimulus (left) and a nasal stimulus (right)
Click here to view |
Children were effectively reinforced for their cooperation and response during the assessment procedure. All the ASP software values and nasometry values were carefully entered into an Excel sheet.
Statistical analysis
Obtained data were subjected to statistical analysis using IBM Statistical Package of the Social Sciences Software for Windows (Version 20. Armonk, New York: IBM Corp). Shapiro–Wilk test of normality revealed normal distribution (P > 0.05) of the data. Mixed ANOVA was done with the “stimulus type” and “instrument type” as within-subject factors and the “participant group” (TDC vs. RCP) as a between-subject factor. Repeated measures ANOVA was done with the “stimulus type” and “instrument type” as within-subject factors. Paired samples t-test was used to compare the nasalance values between the stimuli. Independent samples t-test was used to compare the nasalance values between TDC and RCP, across all the stimuli. Pearson's correlation was carried out to check the correlation between nasalance values from ASP and nasometry.
Results | |  |
The study was a preliminary attempt to investigate and compare the nasalance values of the ASP software with the nasometer in TDC and children with RCP. The nasalance values were compared between the clinical group (RCP) and the control group (TDC), across stimulus type (oral sentences, nasal sentences, and oronasal sentences) and between the instruments (ASP software and nasometer). The mixed ANOVA results revealed that there was a significant interaction effect between: stimulus type and participant group; instrument type and participant group; Stimulus type, instrument type and participant group [Table 2]. | Table 2: Mixed ANOVA results of the interaction effect in between-subject and within-subject factors
Click here to view |
Hence, repeated measures ANOVA was carried out separately for the two participant groups. Within TDC, there was a significant main effect of stimulus type and instrumentation type, and the stimulus type *instrument type interaction was also significant [Table 3]. | Table 3: Repeated measures ANOVA results of the interaction effect in within-subject factors
Click here to view |
Within RCP too, there was a significant main effect of stimulus type and instrument type, but the stimulus type * instrumentation type interaction was not significant [Table 3]. Further results are reported under the following subsections.
Comparison of nasalance values across the stimuli using nasometer and automatic speech processing software in typically developing children
The nasalance values of TDC across the stimuli were derived from the ASP software and nasometer. The mean and standard deviation of these nasalance values is depicted in [Figure 5]. | Figure 5: Mean and standard deviation of nasalance values in percentage from automatic speech processing software and nasometer in typically developing children
Click here to view |
Within TDC, the nasalance values of the nasometer ranged from 15.70% to 51.15%, whereas the nasalance values of the ASP software ranged from 20.65% to 50.25%. In both instrument types, the nasalance values were high for nasal sentences, followed by oronasal sentences and oral sentences. The paired samples t-test results showed that there was a statistically significant difference in the nasalance values across all the stimuli, in both instruments [Table 4]. | Table 4: Paired samples t-test results of the comparison of nasalance values between the stimuli in typically developing children, in both the instrument types
Click here to view |
Comparison of nasalance values across the stimuli using nasometer and automatic speech processing software in children with repaired cleft palate
The nasalance values of children with RCP across the stimuli were derived from the ASP software and nasometer. The mean and standard deviation of these nasalance values is depicted in [Figure 6]. | Figure 6: Mean and standard deviation of nasalance values in percentage from automatic speech processing software and nasometer in children with repaired cleft palate
Click here to view |
Within RCP, the nasalance values of the nasometer ranged from 45.74% to 55.66%, whereas the nasalance values of the ASP software ranged from 57.18% to 67.76%. In both instrument types, the nasalance values were high for nasal sentences, followed by oronasal sentences and oral sentences. The paired samples t-test results showed that there was a statistically significant difference in the nasalance values across all the stimuli, in both instruments [Table 5]. | Table 5: Paired samples t-test results of the comparison of nasalance values between the stimuli in children with repaired cleft palate, in both the instrument types
Click here to view |
Comparison of nasalance values across the stimuli between the groups (typically developing children vs. repaired cleft palate) in both nasometer and automatic speech processing software
In both instrument types, the nasalance values of children with RCP were higher than the nasalance values of TDC in all the stimuli. The independent samples t-test results showed that there was a statistically significant difference in the nasalance values between TDC and RCP, in both the instrument types, across all the stimuli [Table 6]. | Table 6: Independent samples t-test results showing the comparison of nasalance values between typically developing children and repaired cleft palate, in both the instrument types, across all the stimuli
Click here to view |
Comparison of nasalance values between the instruments (nasometer and automatic speech processing software) across the stimuli in both the groups
Within TDC, the nasalance values of ASP were higher than the nasalance values of nasometer in oral sentences and oronasal sentences but not in nasal sentences. The paired samples t-test results showed that, within TDC, there was a statistically significant difference between the nasalance values of ASP and nasometer in oral sentences and oronasal sentences, but not in nasal sentences. In RCP, however, there was a statistically significant difference for nasal and oronasal sentences, but not in oral sentences [Table 7]. | Table 7: Paired samples t-test results of the comparison of nasalance values between automatic speech processing software and nasometer within a different type of stimulus in typically developing children as well as children with repaired cleft palate
Click here to view |
Correlation of nasalance values across the stimuli between nasometer and automatic speech processing software in both the groups
The correlation between the nasalance values of the ASP software and the nasometer was determined separately for oral, nasal, and oronasal sentences, in TDC and children with RCP by administering Pearson's correlation [Table 8]. The results revealed that none of the correlations were significant (P > 0.05). | Table 8: Correlation between the nasalance scores of automatic speech processing software and nasometer in oral sentences, nasal sentences, and oronasal sentences in typically developing children and children with repaired cleft palate
Click here to view |
The correlation results between the nasalance values of nasometry and ASP software showed negative values in children with RCP whereas positive values in TDC, but they were not statistically significant.
Discussion | |  |
The results are discussed under the following subsections.
Comparison of nasalance values across the stimuli using nasometer and automatic speech processing software in typically developing children
Across the stimuli, the ASP software and nasometer followed the same trend, where the nasalance values were high for the nasal sentences followed by the oronasal sentences and the oral sentences. This can be attributed to the physiological reason that the velum remains elevated, and velopharyngeal (VP) closure is maintained for the oral stimuli, whereas the velum lowers and the VP port is opened for the nasal stimuli. The nasalance for the speech segments with combination oral and nasal stimuli falls in between.[2] This finding agrees with the results of the previous nasometer studies.[14],[15],[16],[17] The comparison of nasalance values between each stimulus showed a statistically significant difference in both the instruments. This shows that both the ASP software and nasometer are equally efficient in differentiating oral sentences, nasal sentences, and oronasal sentences using nasalance values. This result ensures that acoustic parameters are useful in identifying hypernasality.
Comparison of nasalance values across the stimuli using nasometer and automatic speech processing software in children with repaired cleft palate
In children with RCP, the nasalance values of the nasometer and ASP software followed the same trend as that in TDC where the nasalance values were high for the nasal sentences followed by the oronasal sentences and the oral sentences. The comparison of nasalance values between each stimulus showed a statistically significant difference in both the instruments. Despite the oral-nasal imbalance due to VP dysfunction in children with RCP, there will be some amount of closure of the VP port for oral stimuli, which depends upon the severity of nasality. This result is in harmony with the results of the previous nasometer studies.[12],[13] This result again ensures the efficiency of the ASP software in identifying nasal resonance.
Comparison of nasalance values across the stimuli between the groups (typically developing children vs. repaired cleft palate) in both nasometer and automatic speech processing software
Independent samples t-test revealed significantly higher nasalance values for children with RCP than for TDC across all the stimuli in both the instruments. This can be attributed to the oral-nasal imbalance due to VP dysfunction in children with RCP. The amplitude of this oral-nasal balance increases with the increased severity of VP impairment.[13],[36] ASP software was successful in identifying nasalance and providing values in TDC and children with RCP. In addition, this result is consistent with previous ASP studies.[20],[21],[22]
Comparison of nasalance values between the instruments (nasometer and automatic speech processing software) across the stimuli in both the groups
Paired samples t-test revealed a significant difference between the nasalance values of ASP and nasometer in oral and oronasal sentences in TDC and nasal sentences and oronasal sentences in RCP. Whereas, such a difference was not found in nasal sentences in TDC and oral sentences in RCP. This discrepancy in the results might be because of the difference in the filter characteristics of the nasometer and ASP software. The upper limit (nasalance for nasal stimuli) of both the instruments is more or less equal, whereas the lower limit (nasalance for oral stimuli) is slightly higher in the ASP software [Table 2] and [Table 3]. This leads to a difference in the dynamic range between the two instruments. The sample size might be an important factor that has led to this difference in the nasalance values across the stimuli between the instruments, especially in RCP.
Correlation of nasalance values across the stimuli between nasometer and automatic speech processing software in both the groups
The nasalance values across the stimuli between nasometer and ASP software in both the groups showed no significant correlations. This shows that the way of nasalance measurement is different in both the instruments. Again, this might be because of the difference in the filter characteristics of the instruments and their dynamic range.
Conclusions | |  |
The present study shows that the ASP software was successful in identifying nasalance in TDC and children with RCP. Like nasometer, the nasalance values of ASP software were significantly different across oral sentences, nasal sentences, and oronasal sentences with nasalance values. However, a major issue needs to be addressed concerning the dynamic range of the software. Future studies should focus on investigating the performance of the software without a sound-treated room. Furthermore, to ensure that the ASP software is an efficient diagnostic tool, it has to be validated on a large number of populations.
Acknowledgment
The authors would like to thank the Director, All India Institute of Speech and Hearing, Mysuru, for providing us the setup to carry out the study. The authors would also like to thank Mr. Srikar Vijayasarathy for doing the statistics and Dr. Amulya. P. Rao for proofreading our research article.
Financial support and sponsorship
Nil.
Conflicts of interest
There are no conflicts of interest.
Appendix | |  |
Appendix 1: Speech Stimuli
- Oral sentences:
/ba: lu ṯabala ba: risu/
/beɖa ka:ɖige oɖiḏa/
/gi:ṯa be: ga ho: gu/
/di: pa da: ri da:ʈidaɭu/
/aḏu ʤo: ga ʤalapa:ṯa/
- Nasal Sentences:
/manu a: nejannu noɖiḏa/
/navi: na manejinḏa banḏanu/
/na: nu a: nejannu noɖiḏe/
/manga maneja me: liḏe/
/ma: ma: manɖjaḏinḏa banḏanu/
- Oronasal Sentences:
/kamala u:ʈa ma:ɖiḏaḏu/
/navi: na pa:ʈa o:ḏiḏa/
/na: nage ʧamaʧa Koɖu/
/beṯṯḏa me: le hoḏalu/
/ra: mana ʤoṯejalli banḏe/
References | |  |
1. | Peterson-Falzone SJ, Trost-Cardamone J, Karnell MP, Hardin-Jones MA. The Clinician's Guide to Treating Cleft Palate Speech-E-Book. 2 nd ed. St. Louis: Mosby Elsevier Health Sciences; 2016. |
2. | Kummer A. Cleft Palate and Craniofacial Anomalies; Effects on Speech and Resonance. 3 rd ed. New York: Cengage Learning; 2014. |
3. | Gutzmann HS. Untersuchungen über das Wesen der Nasalität. Arch Laryngol Rhinol 1913;27:59-125. |
4. | Bzoch KR. Measurement and assessment of categorical aspects of cleft palate language, voice, and speech disorders. In: Communicative Disorders Related to Cleft Lip and Palate. 3 rd ed. Boston: Little, Brown; 1989. p. 137. |
5. | Sell D, Harding A, Grunwell P. GOS.SP.ASS.'98: An assessment for speech disorders associated with cleft palate and/or velopharyngeal dysfunction (revised). Int J Lang Commun Disord 1999;34:17-33. |
6. | Dworkin JP, Marunick MT, Krouse JH. Velopharyngeal dysfunction: Speech characteristics, variable etiologies, evaluation techniques, and differential treatments. Lang Speech Hear Serv Sch 2004;35:333-52. |
7. | John A, Sell D, Sweeney T, Harding-Bell A, Williams A. The cleft audit protocol for speech-augmented: A validated and reliable measure for auditing cleft speech. Cleft Palate Craniofac J 2006;43:272-88. |
8. | Henningsson G, Kuehn DP, Sell D, Sweeney T, Trost-Cardamone JE, Whitehill TL, et al. Universal parameters for reporting speech outcomes in individuals with cleft palate. Cleft Palate Craniofac J 2008;45:1-17. |
9. | Bressmann T. Comparison of nasalance scores obtained with the nasometer, the nasal view, and the Oro nasal system. Cleft Palate-Craniofacial J 2005;42:423-33. |
10. | Fletcher SG. Diagnosing Speech Disorders from Cleft Palate. New York: Grune and Stratton; 1978. p. 92-157. |
11. | Shin YJ, Kim Y, Kim HG. The comparative study of resonance disorders for Vietnamese and Korean cleft palate speakers using nasometer. Maxillofac Plast Reconstr Surg 2017;39:9. |
12. | Vallino-Napoli LD, Montgomery AA. Examination of the standard deviation of mean nasalance scores in subjects with cleft palate: Implications for clinical use. Cleft Palate Craniofac J 1997;34:512-9. |
13. | Navya A. The correlation between derived nasalance measures and perceived nasality in children with repaired cleft lip and palate. Online J Otolaryngol 2014;4:1-12. |
14. | Jayakumar T, Pushpavathi M. Normative value for nasometer in Kannada. In: Master's Thesis. Mysore, KAR: All India Institute of Speech and Hearing; 2005. |
15. | Reddy BM, Sheela S, Kishore Pebbili G. Comparison of Nasalance Values Obtained from Nasality Visualization System and Nasometer II. J All India Inst Speech Hear; 2012;31:1-9.. |
16. | Gnanavel K, Gopisankar R, Pushpavathi M. Normative nasalance values across stimuli and gender in Malayalam speaking individuals. Online J Otolaryngol 2013;3:85-98. |
17. | Chatterjee I, Bhatt SS, Kumari K, Raj D, Saxena V. Influence of Khasi language on nasal and oral passages in English: A nasometric study. Bengal J Otolaryngol Head Neck Surg 2020;28:17-22. |
18. | Niu X. Measurement, analysis, and detection of nasalization in speech. In: Doctoral Dissertation. Oregon, United States: OGI School of Science and Engineering; 2008. p. 28-9. |
19. | van der Heijden P, Hobbel HH, van der Laan BF, Korsten-Meijer AG, Goorhuis-Brouwer SM. Nasometry cooperation in children 4-6 years of age. Int J Pediatr Otorhinolaryngol 2011;75:627-30. |
20. | Orozco-Arroyave JR, Murillo-Rendón S, Álvarez-Meza AM, Arias-Londoño JD, Delgado-Trejos E, Vargas-Bonilla JF, Castellanos-Domínguez CG. Automatic selection of acoustic and non-linear dynamic features in voice signals for hypernasality detection. In: INTERSPEECH 2011 Aug 28-31; Florence, Italy: ISCA (International Speech Communication Association); 2011. p. 529-32. |
21. | Rendón SM, Arroyave JO, Bonilla JV, Londono JA, Domínguez CC. Automatic detection of hypernasality in children. In: Ferrández JM, Álvarez Sánchez JR, de la Paz F, Toledo FJ, editors. International Work-Conference on the Interplay Between Natural and Artificial Computation. Berlin, Heidelberg: Springer; 2011. p. 167-74. |
22. | Vijayalakshmi P,Nagarajan T,Rav J. Selective pole modification-based technique for the analysis and detection of hypernasality. In: TENCON; Singapore: Institute of Electrical and Electronics Engineers (IEEE) 2009:1-5. |
23. | Orozco-Arroyave JR, Vargas-Bonilla JF, Arias-Londoño JD, Murillo-Rendón S, Castellanos-Domínguez G, Garcés JF. Nonlinear dynamics for hypernasality detection in Spanish vowels and words. Cognit Comput 2013;5:448-57. |
24. | Dubey AK, Tripathi A, Prasanna SR, Dandapat S. Detection of hypernasality based on vowel space area. J Acoust Soc Am 2018;143:EL412. |
25. | Golabbakhsh M, Abnavi F, Kadkhodaei Elyaderani M, Derakhshandeh F, Khanlar F, Rong P, et al. Automatic identification of hypernasality in normal and cleft lip and palate patients with acoustic analysis of speech. J Acoust Soc Am 2017;141:929. |
26. | Castellanos G, Daza G, Sánchez L, Castrillón O, Suárez J. Acoustic speech analysis for hypernasality detection in children. Conf Proc IEEE Eng Med Biol Soc 2006;2006:5507-10. |
27. | Maier A, Schuster M, Haderlein T, Nöth E. Hypernasality in speech of children with cleft lip and palate: Automatic evaluation. In: 8 th Int. Seminar on Speech Production. Strasbourg, France: INRIA; 2008. p. 277-80. |
28. | He L, Zhang J, Liu Q, Yin H, Lech M, Huang Y. Automatic evaluation of hypernasality based on a cleft palate speech database. J Med Syst 2015;39:61. |
29. | Prasanna SM, Abraham AK, Dandapat, S, Pushpavathi M, Sathish HV. NASOSPEECH: Development of Diagnostic System for Severity Assessment of the Disordered Speech [Department of Biotechnology Project, Government of India]. Mysuru: IIT, Guwahati and AIISH; 2020. |
30. | Sidhu SS, Deshmukh R. In: Kannapan JG, editor. Cleft Lip and/or Cleft Palate and Orofacial Anomalies: A Multidisciplinary Approach. Chennai: Shanti Anand Printers; 1988. p. 1-18. |
31. | Reddy SG, Reddy RR, Bronkhorst EM, Prasad R, Ettema AM, Sailer HF, et al. Incidence of cleft lip and palate in the state of Andhra Pradesh, South India. Indian J Plast Surg 2010;43:184-9.  [ PUBMED] [Full text] |
32. | Counihan DT, Cullinan WL. Reliability and dispersion of nasality ratings. Cleft Palate J 1970;7:261-70. |
33. | Basavaraj V, Venkatesan S. Ethical Guidelines for Bio-Behavioural Research Involving Human Subjects. Mysore, KAR: All India Institute of Speech and Hearing; 2009. |
34. | Titze IR, Winholtz WS. Effect of microphone type and placement on voice perturbation measurements. J Speech Hear Res 1993;36:1177-90. |
35. | Vikram CM, Tripathi A, Kalita S, Prasanna SM. Estimation of Hypernasality Scores from Cleft Lip and Palate Speech. ISCA (International Speech Communication Association:Interspeech; 2018. p. 1701-5. |
36. | Jones DL. The relationship between temporal aspects of oral-nasal balance and classification of velopharyngeal status in speakers with cleft palate. Cleft Palate Craniofac J 2000;37:363-9.Fugiamusa il |
[Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6]
[Table 1], [Table 2], [Table 3], [Table 4], [Table 5], [Table 6], [Table 7], [Table 8]
|